Christopher W. Goode, Gregory J. Bubalo, BĂşbalo Rotman PLC, Lexington, KY, Gary M. Weiss, J. Allan Cobb, Weiss and Cobb, Leslie M, Cronen, Steven B. Rot-man, BĂşbalo Rotman PLC, Louisville, KY, for Plaintiff., Casey Wood' Hensley, Charles M. Pritchett, Jr., Lori Elizabeth Hammond, Frost Brown Todd LLC, Louisville, KY, Paul S. Penticuff, Baker Sterchi Cowden & Rice, LLC, Kansas City, MO, for Defendant.
Full Opinion (html_with_citations)
MEMORANDUM OPINION AND ORDER
Plaintiffs, Ronna and John Dalton (collectively âthe Daltonsâ), brought this diversity action against Animas Corporation, alleging various tort and contract claims, as well as Kentucky statutory violations. This action arises from injuries Mrs. Dalton suffered when the Animas infusion pump she used somehow dispensed an overdose of insulin. The circumstances of the injurious overdose are disputed. Nevertheless, Defendant has moved for summary judgment on all claims. For the reasons that follow, the motion is sustained in part and denied in part.
I.
Animas Corporation is a Delaware corporation that manufactures external insulin infusion pumps, including the Animas Model 2020 infusion pump (the â2020 Pumpâ). The 2020 Pump automatically injects precise doses of insulin for the treatment of diabetes. Insulin leaves the pumpâs cartridge and is delivered to the user via tubing and a needle attached to the body. The pump software controls the timing and dosage of the infusions. After an insulin cartridge is empty, the user must replace the cartridge, which typically occurs once every three days. When doing so, the user is required to perform a âRewind/Align/Primeâ operation (the âPrime Operationâ). The purpose of the Prime Operation is to remove any air from the system that may result from replacing the insulin cartridge. To perform the Prime Operation, the user must press and physically hold down the pumpâs âOKâ button until insulin is seen dripping from the tube. During this process, the user is warned to always remove the tubing from her body.
Mrs. Dalton used the 2020 Pump without incident for nearly one year. On April 21, 2008, Mrs. Dalton received a low cartridge warning. Although she does not have a specific memory of doing so, Mrs. Dalton believes that she would have replaced the insulin cartridge and performed the Prime Operation. The software confirms this, showing that Mrs. Dalton changed the cartridge and performed the Prime Operation at 10:00 p.m. that evening.
Mrs. Dalton went to bed that night in her bedroom, which is separate from her husbandâs. During the night, Mr. Dalton woke his wife to inform her the Replace Battery Alarm on her 2020 Pump was sounding. The software indicates that the alarm sounded at 4:42 a.m. Mrs. Dalton testified that she got out of bed, went to the kitchen and changed the pumpâs bat
Plaintiffs filed this action against Defendant, claiming that a defect in Mrs. Daltonâs 2020 Pump caused an over-infusion, resulting in personal injuries to Mrs. Dalton. Plaintiffs have asserted claims based on strict liability, negligence, breach of express and implied warranty, violations of the Kentucky Consumer Protection Act, loss of consortium, and punitive damages.
II.
Summary judgement is appropriate where âthe pleadings, depositions, answers to interrogatories, and admissions on file, together with the affidavits, if any, show that there is no genuine issue as to any material fact and that the moving party is entitled to a judgment as a matter of law.â Fed.R.Civ.P. 56(c). Summary judgment is appropriate âagainst a party who fails to make a showing sufficient to establish the existence of an element essential to that partyâs case, and on which that party .will bear the burden of proof at trial.â Celotex Corp. v. Catrett, 477 U.S. 317, 322, 106 S.Ct. 2548, 91 L.Ed.2d 265 (1986). In assessing a summary judgment motion, the court must examine any pleadings, depositions, answers to interrogatories, admissions, and affidavits in the light most favorable to the non-moving party. Fed. R.Civ.P. 56(c).
Plaintiffs brought the following claims: Count I (Strict Liability), Count II (Negligence), Count III (Breach of Express and Implied Warranty), Count IV (Violation of the Kentucky Consumer Protection Act, KRS § 367.170) (âKCPAâ), Count V (Damages), Count VI (Loss of Consortium) and Count VII (Punitive Damage). Several claims alleged in the Complaint warrant dismissal. The Court will address each count individually.
III.
In Kentucky, a plaintiff may prevail on a strict products liability claim when he or she meets the requirements of Restatement (Second) of Torts § 402A, which âimposes strict liability on one who sells any product in a defective condition unreasonably dangerous to the user or consumer.â Leslie v. Cincinnati Sub-Zero Products, Inc., 961 S.W.2d 799, 803 (Ky.Ct.App.1998) (internal citation omitted). Accordingly, a plaintiff has the burden of showing an identifiable, unreasonably dangerous defect. See Gray v. General Motors Corp., 133 F.Supp.2d 530, 533 (E.D.Ky.2001), aff'd, 312 F-3d 240 (6th Cir.2002). Kentucky law is clear that evidence that merely surmises or speculates as to a defect is not sufficient. Midwestern V.W. Corp. v. Ringley, 503 S.W.2d 745, 747 (Ky.1973).
Additionally, the plaintiff must establish causation, that the product was a âsubstantial factorâ in bringing about the alleged harm. See Bailey v. N. Am. Refractories Co., 95 S.W.3d 868, 873 (Ky.Ct.App.2001). Circumstantial evidence may be used to establish legal causation if such evidence is âsufficient to tilt the balance from possibility to probability.â Morales v.. Am. Honda Motor Co., Inc., 151 F.3d 500, 507 (6th Cir.1998).
Since the manufacturing and design defects were both attributed to the defendant, the plaintiff could proceed with two theories of liability against that defendant. The Court reasoned that â[a] plaintiff is not required to show precisely how a product is defective, but simply must show whether it was defective.â Id. A plaintiff can accomplish this by âruling out other theories of causation: â[Wjhere an injury may as reasonably be attributed to a cause that will excuse the defendant as to a cause that will subject it to no liability, no recovery can be had.â â In re Beverly Hills Fire Lit., 695 F.2d 207, 218 (6th Cir.1982) (quoting Suttonâs Admâr v. Louisville & N.R. Co., 168 Ky. 81, 181 S.W. 938, 940 (1916)). In sum, a plaintiff may present two or more potential theories explaining the defect in a products liability action, so long as each theory is attributable to the defendant.
Plaintiffs bring their strict liability claim under three theories: a manufacturing defect (the âstuck button theoryâ) and two design defects (the âprime limit theory,â and the âchange of battery theoryâ). The Court will address each individually.
A.
Plaintiffsâ experts maintain that a manufacturing defect is one possible cause of the over-infusion of insulin. Specifically, according to the âstuck button theory,â the pumpâs âOKâ button stuck during the Prime Operation, causing the device to dispense the entire cartridge of insulin.
Plaintiffsâ experts have had ample opportunity to examine Mrs. Daltonâs pump and have not proffered any evidence of a stuck key. Mr. Michael Klimowicz, Plaintiffsâ electrical engineer expert, manipulated the pump and was unable to recreate the âstuck key theory.â He testified that he pressed the button over thirty times and it never did stick. Absent actual evidence, or circumstantial evidence that rises about mere speculation, that the pump had a defective, stuck button, Plaintiffs cannot pursue their products liability claim under the âstuck button theory.â
B.
Plaintiffs also assert that the 2020 Pump was defectively designed. A product may be defective if it was designed âaccording to an unreasonably dangerous design.â Jones v. Hutchinson Mfg., Inc., 502 S.W.2d 66, 69 (Ky.1973). âThe maker is not required to design the best possible product or one as good as others make' or a better product than the one he has, so long as it is reasonably safe.â Sturm, Ruger & Co. v. Bloyd, 586 S.W.2d 19, 21-22 (Ky.1979). Thus the inquiry in a design defect case is âwhether the product creates such a risk of an accident of the general nature of the one in question that an ordinarily prudent company engaged in the manufacture of such a product would not have put it on the market.â Montgomery Elevator Co. v. McCullough, 676 S.W.2d 776, 780 (Ky.1984) (internal quotation omitted). â[I]n Kentucky, in order to prove a product is âunreasonably dangerousâ as designed, a plaintiff is required to produce competent evidence âof a feasible alternative designâ that would have prevented the injury.â â Cummins v. BIC USA Inc., 835 F.Supp.2d 322, 326 (W.D.Ky.2011) (quoting Toyota Motor Corp. v. Gregory, 136 S.W.3d 35, 42 (Ky. 2004)).
1.
Plaintiffs argue under the âprime limit theory,â that the 2020 Pump was defectively designed because it did not include a prime limit, a feature that limits the amount of insulin dispensed during a Prime Operation. Thus, theoretically, the prime limit feature prevents an over-infusion. Plaintiffsâ experts assert that the prime limit feature would have prevented Mrs. Daltonâs injuries.
Defendant argues that Plaintiffs have failed to prove that the absence of a prime limit caused this incident or that another design would have prevented Mrs. Daltonâs injuries.
The Court finds that Plaintiffsâ prime limit theory establishes that a different design both was feasible and might have prevented Mrs. Daltonâs injuries. As Defendant has argued, users are cautioned to disconnect during the Prime Operation in the 2020 Pump user guide, during user training and in warnings displayed on the pump itself. Mrs. Dalton underwent three hours of training on the use of the 2020 Pump and stated that she knew to disconnect the pump from her body during the Prime Operation.
It is reasonably foreseeable that a user may be in a hypoglycemic state during the Prime Operation and not be cognizant of the risk associated with staying connected to the pump. See Ostendorf v. Clark Equipment Co., 122 S.W.3d 530, 535 (Ky.2003) (stating âit is the legal duty of a manufacturer to use reasonable care to protect against foreseeable dangersâ). Diabetics are prone to experience hypoglycemic episodes. A jury could conclude that Defendantâs warnings and training were insufficient to make the 2020 Pump safe for its intended users. Therefore, Defendant should have anticipated misuse of their product given that such a userâs memory and cognitive ability may, at times, be impaired. A prime limit would prevent an impaired or incoherent user from over-infusing oneself when inadvertently attached during the Prime Operation. A reasonable jury could find, therefore, that Defendant should have anticipated such misuse and designed the device with a prime limit feature, and that the 2020 Pump was unreasonably dangerous for lacking such a feature.
2.
Plaintiffs additionally argue that the 2020 Pump is defective because it requires a Prime Operation following a battery change, thus exposing users to a medically unnecessary Prime Operation. Defendant counters that a Prime Operation is necessary following a battery change to reinitiate the force sensor and cartridge position and ensure no conditions have changed
The âchange of battery theoryâ is related to the âprime limit theory,â in that a prime limit would have prevented an over-infusion during a Prime Operation initiated by a change of battery. However, this theory is also independent from the âprime limit theory.â Plaintiffs have proffered evidence of other pumps on the market that did not require a prime after a battery change. The prime that resulted in Mrs. Daltonâs over-infusion would not have occurred had the 2020 Pump not required a Prime Operation following a battery change. Defendant argues that requiring a Prime Operation in this instance protected against a greater risk, the chance that the device misaligns during a change of battery and introduces air into the system. Whether the Prime Operation is necessary and, correspondingly, whether the device is unreasonably dangerous by requiring this process following a battery change, is sufficiently disputed for a jury to decide.
Accordingly, the Court finds summary judgment on Plaintiffsâ design defect claim is inappropriate at this juncture.
IV.
âA plaintiff in Kentucky can bring a defective design claim under either a theory of negligence or strict liability.â Ostendorf, 122 S.W.3d at 535. Both Plaintiffsâ strict liability and negligence claims are premised on their argument that the product is unreasonably dangerous, ie., that the product created an unreasonable risk of foreseeable injury. However, negligence focuses on the conduct of the manufacturer, and specifically whether the manufacturer used reasonable care to protect against foreseeable dangers, while strict products liability focuses on the defective product itself. Id. Courts employ a risk-utility balancing test that assesses the manufacturerâs decision to design a product in a certain manner and whether the manufacturer exercised reasonable care in making the design choices it made. Id.
Because the Court finds that a reasonable jury could find that the 2020 Pump had a design defect, a reasonable jury could also find that Animas was negligent in its design of the 2020 Pump. See Ostendorf, 122 S.W.3d at 535 (holding that a plaintiff may bring a defective design claim under a theory of negligence or strict liability and both theories are premised on the argument the product is unreasonably dangerous); see also Nichols v. Union Underwear Co., Inc., 602 S.W.2d 429, 433 (Ky.1980) (stating the distinction between negligence and strict liability âis of no practical significance so far as the standard of conduct required of the defendant ... is reasonable careâ). Accordingly, the negligence cause of action remains. However, as noted above, a defective design claim proceeds under a theory of negligence or strict liability. At a recent conference in chambers, counsel for Plaintiffs advised the Court that they planned to primarily pursue the strict liability claim. Though it may be possible for Plaintiffs to pursue different theories for recovery in a design defect case, the Court cautions a negligence claim duplicating the recovery available under strict liability generally does not add anything to the case and risks jury confusion.
V.
Plaintiffsâ claims for both breach of warranty and for violation of the KCPA seem likely to fail and do not add substantial value to their case.
If a product is proven to be defective, a Defendant may be contractually liable for a breach of warranty. McCoy v.
Plaintiffsâ claim for breach of an implied warranty of fitness for a particular purpose fails because Mrs. Daltonâs use of the 2020 Pump was not peculiar. See KRS § 355.2-315 (âWhere the seller at the time of contracting has reason to know any particular purpose for which the goods are required ... there is ... an implied warranty that the goods shall be fit for such purpose.â). Mrs. Dalton used the 2020 Pump for its ordinary purpose â as an insulin delivery product. Absent an allegation that Mrs. Dalton used the device for a particular purpose of which Defendant would have reason to know, this claim must be dismissed.
The Plaintiffsâ claim based on alleged violations of the KCPA must also be dismissed. The KCPA prohibits â[u]nfair, false, misleading, or deceptive acts or practices in the conduct of any trade or commerce,â where âunfair shall be construed to mean unconscionable.â KRS § 367.170. Plaintiffs have not identified any conduct of Defendant that constituted a KCPA violation. Rather, their Complaint generally states, âAnimas misrepresented the efficacy and safety of the Animas 2020 insulin infusion pump.â The Plaintiffsâ Response to Defendantâs Motion for Summary Judgment does not defend their KCPA claim.
The Courtâs review of the record has not revealed a viable claim. Plaintiffs have not alleged facts upon which a reasonable jury could find Defendant intentionally, knowingly, or in bad faith took malign actions affecting Plaintiffs. Therefore, Plaintiffsâ KCPA claim fails as a matter of law.
VI.
In Kentucky, a claim for âloss of consortium is derivative of an injured plaintiffs claim.â Norton v. Canadian Am. Tank Lines, 2009 WL 931137, at *2 (W.D.Ky. Apr. 3, 2009) (citing Daley v. Reed, 87 S.W.3d 247 (Ky.2002)). As such, Mr. Daltonâs claim is entirely dependent upon the success of his wifeâs product liability claim. Given that Plaintiffs do have remaining claims against Defendant, summary judgment on Mr. Daltonâs loss of consortium claim is inappropriate at this time.
During a pre-trial conference with both parties, it was clear to the Court that the Plaintiffsâ main objective in trial is to pursue the products liability claim based on a design defect in the 2020 Pump. The negligence, breach of warranty, and loss of consortium claims appear to be secondary. Most likely, they will be accounted for if and when the jury is charged with assessing liability and damages.
VII.
Count V of the Complaint, entitled âDamages,â asserts a nebulous cause of action for various types of damages. This count is dismissed as a matter of law, because damages are a prayer for relief, not a cause of action.
Defendant has also moved for summary judgment as to Plaintiffsâ claim for punitive damages, which is set forth as a separate cause of action in Count VII. However, a claim for punitive damages is not a separate cause of action, but a remedy potentially available for another cause of action. Toon v. City of Hopkinsville,
To proceed on a claim for punitive damages Plaintiffs must have some evidence from which a reasonable jury could find malice or reckless disregard for a product userâs safety. Kinney v. Butcher, 131 S.W.3d 357, 359 (Ky.Ct.App.2004). Here, the evidence is that Defendant took many actions to ensure user safety. The question is whether those actions were reasonably sufficient. Thus, while the evidence may raise issues of negligence, it does not suggest any questions as to malice or malicious disregard. The evidence does not submit to such an inference. Consequently, Plaintiffsâ claim for punitive damages must be dismissed.
Being otherwise sufficiently advised,
IT IS HEREBY ORDERED that Defendantâs Motion for Summary Judgment is GRANTED in part and Plaintiffsâ claims for a manufacturing defect (the âstuck button theoryâ), under the KCPA and for punitive damages are DISMISSED WITH PREJUDICE.
The remaining claims are Count I (Strict Liability with respect to the defective design theories), Count II (Negligence), Count III (Breach of Express Warranty) and Count VI (Loss of Consortium).
. The law distinguishes between two types of defects: (1) a manufacturing defect, which is a deviation from a product's design that creates an unreasonable risk of harm, Wright v. General Elec. Co., 242 S.W.3d 674, 682 (Ky.Ct.App.2007); and (2) a design defect, which is a defect that âexists when a product is built in accordance with its intended specifications, but the design itself is inherently defective or poses unreasonable dangers to consumers.â 63A Am. Jur. 2d Products Liability § 869 (2012).
. Citing the recent Siegel case, Defendant argues that Plaintiffs may not use an expert's opinion to infer causation unless Plaintiffs eliminate all other reasonable explanations for the accident. However, this opinion is consistent with the recent Siegel case. There are no proffered alternative explanations for the over-infusion that would absolve Defendant of its liability. Defendant maintains that an alternative cause for Mrs. Daltonâs injury is that she did not follow known warnings by failing to disconnect from the pump during the Prime Operation. However, in that circumstance, a prime limit would have prevented the over-infusion. As will be discussed, a reasonable jury could find the 2020 Pump to be unreasonably dangerous despite the directives to disconnect the pump during the Prime Operation because it lacked a prime limit feature.
. Defendant argues that the Plaintiffsâ expertsâ inability to determine what the prime limit should have been renders this theory irrelevant. Plaintiffsâ experts do not agree or disagree on what the prime limit should be, each suggesting a limit somewhere between 5 and 30 units of insulin. Defendant contends that a 30-unit prime limit would likely have resulted in injuries to Mrs. Dalton; thus a prime limit feature would not have prevented
. The Court notes that Mrs. Dalton has received previous training on a different insulin pump manufactured by Medtronic, but these three hours focused primarily on features in Animas' 2020 Pump.